This banner reel contains paid advertising

Equipment and Calibration Requirements

The following equipment is the minimum required to undertake testing under the SITMA Scheme. Testers must ensure they have access to all items relevant to the test type they are performing.

Test Type Item
Airborne (Part 1)
Sound Level Meter
Speaker ( & Amplifier)
Calibrator
Impact (Part 2)
Sound Level Meter
Speaker & Amplifer
Calibrator
Tapping Machine

Calibration of Measurement Equipment

Accredited Calibration

Calibration certificates from manufacturers, distributors or non-accredited laboratories are not acceptable for items requiring accredited calibration.

Certificates must be issued by a laboratory accredited to ISO/IEC 17025:2017 by UKAS, or by a national accreditation body recognised through an appropriate international mutual recognition arrangement. The laboratory’s accredited scope shall include the relevant measurement activity, measurement range and applicable calibration standard.

It is noted that calibration of older devices can be completed to newer standards than identified below, but it is not acceptable to measure a newer device to an older standard.

Precision Requirements (SLM)

Sound level meters shall measure and report sound pressure level to a minimum resolution of 0.1dB (0.1 decimal place)
Sound level meters shall measure and report reverberation time (RT) to a minimum resolution of 0.01s (2 decimal places).

Manufactured Standard vs Calibration Standard

The following calibration standards are required by the SITMA Certification Scheme

Sound Level Meter Conformance Calibration Requirement
IEC 60651 & IEC 60804 (pre-IEC SLMs)
BS 7580 Part 1:1997
IEC 61672-1:2002
IEC 61672-3: 2006
IEC 61672-1:2013
IEC 61672-3:2013

Filter Calibration

The following calibration standards are required by the SITMA Certification Scheme

Sound Level Meter Conformance Calibration Requirement
IEC 60651 & IEC 60804 (pre-IEC SLMs)
IEC 61260:1995 (or BS EN IEC 61260:1996)
IEC 61672-1:2002
IEC 61260:1995 (or BS EN IEC 61260:1996)
IEC 61672-1:2013
IEC 61260-3:2016 (or BS EN 61260-3:2016) 3.5

Reverberation Calibration

The following calibration standards are required by the SITMA Certification Scheme

Sound Level Meter Conformance Calibration Requirement
IEC 60651 & IEC 60804 (pre-IEC SLMs)
Suitable to support the verification of specific RT modules on sound level meters using transfer reference audio files i.e. computer generated multi-sine files to give the required decay curves.

IEC 61672-1:2002 and IEC 61672-1:2013 devices must demonstrate an maximum expanded measurement uncertainty of <k = 2 in each frequency

Pre-IEC SLMs must still be verified and meet the above requirements, though it is accepted by the SITMA scheme that there are limitations to both measuring at each 1/3 Octave frequency bands and short measurement durations (0.3s).

Where measurements cannot be recorded within acceptable ranges, a permanent deviation note will be applied on to each test report.
IEC 61672-1:2002
IEC 61672-1:2013

Acceptance of Equipment

Acceptance of calibration does not constitute automatic acceptance of the instrument for use within the SITMA Scheme. The tester shall provide sufficient
evidence to demonstrate:
• the make, model and serial number of the instrument;
• the standard against which the instrument model was originally type or pattern evaluated;
• a valid accredited calibration certificate against that standard;
• suitable filter performance across the frequencies used for SITMA testing;
• suitable reverberation-time measurement capability;
• the ability to measure and report the required one-third octave bands;
• compliance with SITMA’s minimum resolution requirements; and
• compatibility with the SITMA Portal.

Where the evidence is incomplete, inconsistent or does not cover the functions used for SITMA testing, SITMA may require additional verification or determine that the instrument is not suitable for use within the Scheme.

The acceptance decision and supporting evidence shall be recorded by SITMA.

Equipment Requirements

Item Accuracy Calibration Requirements
Sound Level Meter (SLM)
The sound level measurement equipment shall meet the Class 1 or Type 1 requirements of the standard against which the instrument model was originally type or pattern evaluated.

The calibration certificate shall identify the standard used and confirm that the instrument meets the applicable Class 1 or Type 1 requirements. The calibration standard used shall be consistent with the instrument’s original type or pattern evaluation.

Calibration against an earlier standard does not, by itself, demonstrate compliance with SITMA’s separate requirements for octave-band filters, reverberation-time measurement, functional capability or data-export compatibility.

QMS037 provides further justification for the recognition of the applicable sound level meter standards.
Calibration shall be undertaken at least every 24 months, or following any damage, repair or modification that may affect measurement performance.

Calibration shall be completed by a laboratory accredited to ISO/IEC 17025:2017 by UKAS, or by another national accreditation body recognised through an appropriate international mutual recognition arrangement. The laboratory’s accredited scope shall include the relevant calibration activity and applicable calibration standard.

For sound level meters originally type or pattern evaluated against BS EN 61672-1:2013, calibration shall be undertaken in accordance with the applicable requirements of BS EN 61672-3:2013.

For legacy sound level meters originally type or pattern evaluated against an earlier standard, calibration against that earlier standard is acceptable where:
• evidence identifies the standard against which the model was originally type or pattern evaluated;
• an appropriate pattern evaluation is not available to support calibration to BS EN 61672-3:2013;
• the calibration is within the accredited scope of the calibration laboratory; and
• the instrument separately meets SITMA’s applicable filter, reverberation time, functional and data-export requirements.

The certificate and supporting evidence shall demonstrate that the calibration standard is appropriate to the instrument model.
Filters*
Filters shall comply with the Class 1 requirements of BS EN 61260:1996, BS EN 61260-1 or the applicable standard under which the filter set was originally type or pattern evaluated.

Where the sound level meter or filter set cannot be calibrated against the current standard because appropriate

type or pattern evaluation information is unavailable, verification against the applicable earlier standard may be accepted. The calibration certificate shall identify the frequencies assessed and confirm the applicable Class 1 performance.

The filter verification shall cover all one-third octave bands required for SITMA sound insulation testing.
*Calibration shall be undertaken at least every 24 months, or following any damage, repair or modification that may affect measurement performance.
Reverberation Time*
The reverberation-time measurement function shall beThe reverberation-time measurement function shall be verified at the one-third octave band frequencies required for SITMA sound insulation testing between 100 Hz and 3150 Hz.

For instruments capable of being verified at each individual one-third octave band, verification shall cover every one-third octave band between 100 Hz and 3150 Hz.

For legacy instruments whose reverberation-time function cannot be verified using the current method because of the instrument’s original design, type or pattern evaluation, an earlier technically appropriate verification method may be accepted.

This is subject to all of the following:
• the sound level meter was originally type or pattern evaluated against an earlier standard;
• the calibration or verification laboratory confirms that the current method cannot appropriately be applied to that instrument;
• the earlier method is identified on the calibration or verification certificate;
• the laboratory’s accredited scope covers the activity stated on the certificate;
• the instrument remains capable of measuring and reporting the one-third octave bands required by the SITMA test method; and
• SITMA has completed and recorded an equipment compatibility review before accepting the instrument.

A certificate covering only a limited selection of frequencies shall not automatically demonstrate that the instrument’s reverberation-time function is suitable across the full SITMA measurement range. SITMA shall consider the instrument’s original approval standard, technical capability and the verification method used before determining whether the evidence is acceptable.
Calibrator
The accuracy of the sound calibrator must comply and be verified against the requirements of accuracy class 1 defined in IEC 60942:2017.

The calibrator must produce a noise level of either 94.0dB or 114.0dB at a frequency of 1000Hz.
Laboratory accredited to ISO/IEC 17025:2017 by UKAS, or by another national accreditation body whose recognition is accepted by SITMA, for the relevant calibration activity; calibrated every 12 months or after any damage or repair.
Speaker and Tripod
The speaker and noise source:
• shall be capable of producing at least 90.0dB at all one-third octave band frequencies (100 – 3150Hz) in normal use (reverberant room).
• Requirements of PUS021 must be met regarding speaker positioning and stability requirements
To be assessed and checked at the Initial Audit

Tester shall verify for each test they do
Stability Check
The speaker and noise source shall:
• produce a steady, flat spectrum (ensures different parallel frequencies are no more than 6dB different).
Completed either by the tester, or by a 3rd party verification company on entry every 2 years.

Must be documented on the SITMA Portal

(PUS021 – Speaker Stability and Directivity Checks, and/or the section below this table)
Directivity Check
• produce a continuous stable noise.
Completed either by the tester, or by a 3rd party verification company on entry every 2 years.

Must be documented on the SITMA Portal.

(PUS021 – Speaker Stability and Directivity Checks, and/or the section below this table)
Those who wish to be in part 2 of the scheme must also comply with the below
Tapping Machine
The Tapping Machine shall comply and be verified with the details as discussed in Annex A of ISO 140-7:1998
Laboratory accredited to ISO/IEC 17025:2017 by UKAS, or by another national accreditation body whose recognition is accepted by SITMA, for the relevant calibration activity; calibrated every 24 months or after any damage or repair.

Drop heights checked before each test.
BS EN ISO-7:1998 Appendix A “one off checks”
Completed either by the tester, or by a 3rd party verification company when registered on the portal.

Must be documented on the SITMA Portal

(See section 4.3 – Tapping Machine one off checks)

Additional Requirements

Speaker Stability and Speaker Directivity Checks

Checks can be completed by an external laboratory whom themselves hold UKAS accreditation but are not specifically accredited to conduct the speaker checks.

Where these checks are completed externally, and the laboratory holds UKAS accreditation, the FOS016 and FOS017 are not required to be completed.

A certificate of completion is acceptable to SITMA on the basis that it contains the minimum information as follows:

Your How To Complete Guides for Speaker Stability and Directivity Checks

Tapping Machine One-Off Checks

BS EN ISO 140‑7:1998 Annex A defines the mechanical and performance requirements for a standard tapping machine. The parameters listed in this section shall be confirmed before a tapping machine is authorised for use within the SITMA Scheme.

Periodic calibration conducted by an accredited laboratory may cover some, but not all, of the Annex A requirements. Where any required parameter is not included in the calibration report, the tester shall measure and record the values using the SITMA Lodgement Portal form ‘FOS022 – Tapping Machine One‑Off Check Template’. This can be requested by emailing SITMA.

The following parameters are required to be measured once only, unless the tapping machine has been modified, repaired, or had components replaced. These constitute the “one‑off checks” required under ISO 140‑7 Annex A.

SITMA Members can use our template to enter their data in and submit their “one off” Tapping Machine Checks. The template can be found on the SITMA Portal.

Our current issue is: Issue 1.1 – 29th April 2024

One Off Check Requirement Discussion
Distance between hammers
100mm ±3mm
The tapping machine shall have five hammers placed in a line. The distance between centrelines of neighbouring hammers shall be (100 ± 3) mm.
Supports of the tapping machine
>100mm
The distance between the centre of the supports of the tapping machine and the centrelines of neighbouring hammers shall be at least 100 mm. The supports shall be equipped with vibrational insulating pads.
Time between impact & lift
80ms
Can be self recorded using a slow motion phone camera. Figure to be declared.
Maximum possible fall height
No specific limit
No strict requirement. Maximum fall height to be declared. Feet to be raised to a maximum height and maximum drop height measured.

Software, Firmware, Hardware & Data Compatibility

Manufacturers Software/Firmware

SITMA does not assess, verify, or approve manufacturer software, firmware or hardware. Manufacturers declare compliance with the relevant acoustic standards (e.g., BS EN 61672‑1:2013), and testers remain responsible for ensuring their equipment and associated software remain compliant through valid accredited calibration and correct configuration.

Data-Export Compatibility

SITMA’s review is limited solely to confirming that the file produced by the tester’s equipment or software can be correctly processed by the SITMA Portal. This is a compatibility check only; it is not a verification of measurement accuracy, processing algorithms, or software functionality.

File‑format compatibility forms one of the equipment acceptance criteria defined in Section 6 – Equipment Compatibility.

Equipment Compatibility

SITMA does not approve, certify, or validate sound level meters or other test equipment. Equipment is accepted for use only where an objective verification review confirms that the model and the individual instrument meet all of the following criteria:

Manufacture Declaration
The manufacturer specifies the instrument as compliant with the appropriate standard(s), e.g. BS EN 61672‑1:2013 Class 1.
Accredited Calibration Evidence
The individual instrument holds a valid accredited calibration certificate confirming:
Class 1 performance against BS EN 61672-1:2013, where the instrument model was type or pattern evaluated against that standard; or
Class 1 or Type 1 performance against the applicable earlier standard, where the instrument was originally type or pattern evaluated against that earlier standard and calibration to BS EN 61672-3:2013 is not technically appropriate;
Class 1 filter performance against BS EN 61260 or the applicable earlier filter standard;
reverberation-time verification appropriate to the instrument’s design and original type or pattern evaluation;
the frequencies and functions covered by the calibration or verification; and
any limitations or exclusions applying to the calibration evidence.

Where an earlier calibration or verification method has been used, SITMA shall review the evidence to confirm that the instrument remains technically suitable for the measurements required by the Scheme
Data‑Export Compatibility
The instrument can produce measurement files in a format compatible with the SITMA Portal for audit and verification purposes.
Functional Capability
Where SITMA measurements require it, the instrument must support the necessary building acoustics/RT measurement functions

A list of sound level meters currently known to be compatible with the SITMA Portal and to also comply with the requirements of 6.1.1.

MakeModelUpload TypeUK DistributorBuilding Acoustics Module Mandatory?
Brüel & Kjær2250.rndBruel & Kjaer
www.bksv.com/
Yes
Brüel & Kjær2270.rndBruel & Kjaer
www.bksv.com/
Yes
NorsonicNor-140.xslxANV Measurement Systems
www.anv.uk.comk/
Yes
NorsonicNor-145.xslxANV Measurement Systems
www.anv.uk.comk/
Yes
NorsonicNor-150.xslxANV Measurement Systems
www.anv.uk.comk/
Yes
NorsonicNor-118.xslxANV Measurement Systems
www.anv.uk.comk/
No (workaround required)
RionNA-28.xlsxANV Measurements Systems
www.noise-and-vibration.co.uk/
No
RionNA-58.xlsxANV Measurements Systems
www.noise-and-vibration.co.uk/
No
NTi AudioXL2.txtNTi Audio
www.nti-audio.com
Yes
NTi AudioXL3.txtNTi Audio
www.nti-audio.com
Yes
Svantek979.sprSvantek
www.acsoft.co.uk/svantek
No (app required)
Svantek977.sprSvantek
www.acsoft.co.uk/svantek
No (app required)
Svantek971A.sprSvantek
www.acsoft.co.uk/svantek
No (app required)

Ref: PUS007 – Issue 3.1